Surviving Chromatography - Part 2: Corrective Action
Prezentace | 2018 | Agilent TechnologiesInstrumentace
HPLC
ZaměřeníVýrobceAgilent Technologies
Klíčová slovacolumn, plugged, pressure, frit, mobile, replace, phase, shape, causes, problems, solvent, peak, detach, peaks, ecv, retention, issues, filter, volume, restricted, baseline, cause, sample, shifts, problem, beginner, buffer, high, still, cell, lamp, correct, inlet, pitfalls, agilent, possible, check, flush, system, contaminated, corrective, changes, flow, noisy, selectivity, change, can, blockage, always, seat
Podobná PDF
The Chromatography Detective: Troubleshooting Tips & Tools for LC & LCMS
2018|Agilent Technologies|Prezentace
The Chromatography Detective: Troubleshooting Tips & Tools for LC & LCMS Rita Steed LC Columns Application Engineer January 31, 2018 What do you do when..... ➢ Your chromatography changes • Peak shape • Retention ➢ You can’t reproduce a method…
Klíčová slova
beginner, beginnerpitfalls, pitfallsavoiding, avoidingcolumn, columndwell, dwellvolume, volumemobile, mobilephase, phaseecv, ecvpeak, peakinstrument, instrumentshape, shapetailing, tailingconnections, connectionsretention
Taking the Trouble Out of TroubleshootingIs It the Column, Method, or Instrument
2016|Agilent Technologies|Prezentace
Taking the Trouble Out of Troubleshooting Is It the Column, Method, or Instrument Rita Steed September 29, 2016 Page 1 What Do We Troubleshoot Typical LC troubleshooting approach asks: – What’s wrong with my column? – What’s wrong with my…
Klíčová slova
column, columnmobile, mobilephase, phaseproblems, problemsfrit, fritmau, maupeaks, peakstailing, tailingplugged, pluggedfittings, fittingspressure, pressurepeak, peakmin, minshape, shapevolume
HPLC Column Troubleshooting: Is It Really The Column?
2010|Agilent Technologies|Prezentace
HPLC Column Troubleshooting: Is It Really The Column? Agilent Technologies Technologies, Inc. Inc Rita Steed Application Engineer J January 22 22, 2010 Page 1 Agilent Restricted mA U Troubleshooting in HPLC 20 00 15 00 10 5 00 0 0…
Klíčová slova
restricted, restrictedagilent, agilentcolumn, columnmin, minmobile, mobiletailing, tailingtime, timephase, phasechange, changeretention, retentionplugged, pluggedplates, plateslot, lotchanges, changespeak
LC Column Troubleshooting - Isolating the Source of the Problem
2013|Agilent Technologies|Prezentace
LC Column Troubleshooting Isolating the Source of the Problem Rita Steed December 12, 2013 Page 1 What Do We Troubleshoot The typical LC troubleshooting approach asks the questions: – What’s wrong with the column? – What’s wrong with the instrument?…
Klíčová slova
column, columntailing, tailingmin, minmau, mauvolume, volumemobile, mobilefittings, fittingspeak, peakphase, phaseplugged, pluggedpeaks, peakstime, timefrit, fritred, redextra